For units with a Microprocessor version # ending in WP (the version # is displayed on the screen each time the unit is turned on).
Place unit on a sturdy level counter.
Fill the Reservoir with distilled water
The unit is ready to operate
To Adjust the Filling Time:
Press the STOP Key repeatedly until the message “Code:xxx” appears.
Using the UP/DN arrow keys change the code to 105, then press the STOP Key.
A message will be displayed saying “Water in = xx sec”
Using the UP/DN arrow keys change the seconds according to the following:
EZ9 = 30 sec
EZ10 = 35 sec
3870EK = 65 sec
Now press the STOP Key
On some older units the message “Ea Type:” will appear,using the UP/DN arrow keys select either “0” for an E or EK type unit or “1” for an EA or EKA type unit.
Press the STOP Key to finish
The unit is ready to operate
For units with a Microprocessor version # NOT ending in WP
Follow steps 1–5 for Manual Units above,then empty the water from the chamber and check the automatic water fill.
Turn the unit on
With the door open press and hold the door switch,then press the start button. When water starts flowing release the door switch.
Water should come up to the same spot as the measured amount had.
If need be adjust the automatic fill using the following procedure.
To Adjust the Automatic Filling:
While leaving the door open turn power off.
Press and hold the Manual Water Fill Button (this is the button with the two arrows).
Turn the power back on
When the normal display screen appears release the fill button –immediately press it in again and monitor the water flow
Continue to hold the button until water reaches groove at front –then release button – wait ten seconds – the unit is reprogrammed.